B.11 Bruker Pulse Programmer EP385

Module name: ep385

Description: Bruker Pulse Programmer EP385, used as pulser for the Berlin X-band spectrometer.

Timebase: internal clock 8 ns, external clock 8 ns and slower
Minimum repetition time: internal clock 1.286 ms, external clock 160768 times timebase
Repetition time increments: internal clock 102.4 us, external clock 12800 times timebase
Maximum number of pulses per channel: 59
Output channels: 8

Controlled via: GPIB bus (IEEE 488).
In the GPIB configuration file use "EP385" as the device name, '\r' or 0xd as the EOS character, set set-reos, set-xeos to yes and set-bin to no.

Status: Tested

Supported functions:

`pulser_name()'
`pulser_state()'
`pulser_update()'
`pulser_shift()'
`pulser_increment()'
`pulser_reset()'
`pulser_next_phase()'
`pulser_phase_reset()'
`pulser_pulse_reset()'
`pulser_shape_to_defense_minimum_distance()'
`pulser_defense_to_shape_minimum_distance()'
`pulser_automatic_shape_pulses()'
`pulser_automatic_twt_pulses()'
`pulser_minimum_twt_pulse_distance()'
`pulser_show_pulses()'
`pulser_dump_pulses()'

Back: B.10 EIP Microwave Counter 371 Forward: B.12 EPR Modulation Module   FastBack: B. Device Reference Up: B. Device Reference FastForward: C. Reserved Words

This document was generated by Jens Thoms Toerring on September 6, 2017 using texi2html 1.82.