B.22 Tektronix Stimulus System HFS9000 (Berlin W-Band)
Module name: hfs9000
Description: Tektronix Stimulus System HFS9000, used as pulser in the Berlin W-band spectrometer. No support for phase cycled experiments.
Timebase: between 1.6 ns and 20 us
Maximum pulse voltage: between -1.5 V and 5.5 V
Minimum pulse voltage: between -2 V and 5 V
Difference must be not more than 5.5 V and at least 0.5 V
Resolution of pulse voltages: 10 mV
Trigger-in levels: between -4.7 V to 4.7 V
Resolution of trigger-in voltages: 10 mV
Number of channels: 4 plus a TRIGGER_OUT channel
Fixed length of TRIGGER_OUT pulse: 20 ns
Controlled via: GPIB bus (IEEE 488).
In the GPIB configuration file use "HFS9000"
as the device name,
'\n'
or 0xa
as the EOS character, set set-reos
and
set-xeos
to no
and set-bin
to yes
.
Status: Tested
Supported functions:
- `pulser_name()'
- `pulser_state()'
- `pulser_channel_state()'
- `pulser_update()'
- `pulser_shift()'
- `pulser_increment()'
- `pulser_reset()'
- `pulser_pulse_reset()'
- `pulser_lock_keyboard()'
- `pulser_stop_on_update()'
- `pulser_keep_all_pulses()'
- `pulser_maximum_pattern_length()'
- `pulser_show_pulses()'
- `pulser_dump_pulses()'
![]() |
![]() |
![]() |
![]() |
![]() |
This document was generated by Jens Thoms Toerring on September 6, 2017 using texi2html 1.82.